SEM Vibration Isolation #33
Labels
No labels
Completed
Needs Documentation
Stale
Type
Development
Type
Equipment
Type
Process
Type
Research
No milestone
No project
No assignees
2 participants
Notifications
Due date
No due date set.
Blocks
Depends on
#32 SEM Imaging@10nm
fafo/techtree
#28 Scanning Electron Microscope (SEM)
fafo/techtree
#40 Air Compressor
fafo/techtree
Reference: fafo/techtree#33
Loading…
Reference in a new issue
No description provided.
Delete branch "%!s()"
Deleting a branch is permanent. Although the deleted branch may continue to exist for a short time before it actually gets removed, it CANNOT be undone in most cases. Continue?
Our SEM "Kevin" is currently missing its active vibration isolation.
Partial Techtree
Digest: 4edd31396a34eb1d8bffef41580cf29c1bae08ee3912a3a11841a961822d423b; Last Updated: 2025-06-15 00:29:07
@q3k, @0ff: At what imaging scale do we currently hit the vibration limit? Can you add an appropriate dependency from the respective issue to this one here?
For reference, here is an image where vibration artefacts are very visible:
Bought some components so we can at least fill up the pneumatic springs so the SEM column is no longer directly coupled to the building. It is unclear what kind of active vibration isolation system was installed previously, but the springs look like a passive use might be possible.
Will report back once this has been tested.