Rahix
152290f5a3
The accesses to -KEC1 always time out on the attempt to update the output values. We noticed that this is related to the timing between the reading of the inputs and the following write to the outputs. Fix -KEC1 accesses by waiting before sending the next request to the board after receiving the reply for the previous one. |
||
---|---|---|
.. | ||
succd | ||
panel.pdf | ||
panel.qet | ||
README.md | ||
relayboard-din-mount.FCStd | ||
succbone-control-panel.jpg | ||
succbone-din-mount.FCStd | ||
succbone-din-mount.stl |
succbone
succbone is our "automation" system for the vacuum system of the SEM.
Features
- Read a TPR280 pirani gauge (0-10V analog input)
- 8x relay outputs for
- Roughing pump interlock
- Diffusion pump interlock
- Emulate Pump-Down push button
- Emulate Vent push button
- Signal two different vacuum levels to the original evacuation controller
- Rough vacuum reached
- High vacuum reached