succd
A little daemon for monitoring the SEM. This is a temporary solution that runs on a BeagleBone Enhanced (succbone.lab) and SHOULD be replaced with a proper PLC/SCADA system and general process control in the future.
Features:
- Monitors the state of the Pirani gauge via the BBE's builtin ADC.
- Allows enabling/disabling the diffusion/roughing pumps (builtin controller always keeps them enabled).
- Allows for simulating vent/pumpdown button presses.
Pins
| P9 pin number | Relay number | Relay board / channel number | GPIO number | Function | 
|---|---|---|---|---|
| 27 | -KFA1 | 1 / 4 | 115 | Roughing Pump | 
| 23 | -KFA2 | 1 / 3 | 49 | Diffusion Pump | 
| n/c | -KFA3 (broken) | 1 / 2 | n/c | |
| 12 | -KFA4 | 1 / 1 | 60 | Vent Button | 
| 15 | -KFA5 | 2 / 4 | 48 | Pump Down Button | 
| 11 | -KFA6 | 2 / 3 | 30 | Rough Vacuum Out | 
| 42 | -KFA7 | 2 / 2 | 7 | High Vacuum Out | 
| n/c | -KFA8 | 2 / 1 | n/c | 
Accessing at the lab
Go to succbone.lab.fa-fo.de.
Known issues
Sometimes the websocket doesn't connect. Refreshing the page a few times should fix it.
Running locally
$ go run . -fake
Then point your browser to localhost:8080
Deploying on the succbone
$ ssh root@succbone systemctl stop succd
$ GOARCH=arm go build .
$ scp succd root@succbone:/usr/bin/succd