succd: do not early return on error in modbusUpdate
When one device fails, this should not influence updates of the other devices. Thus, early return was the wrong strategy here. Instead, when communication with a device fails, skip the process data update and continue with the next device.
This commit is contained in:
parent
d8a467a0c4
commit
6f93b96c39
|
@ -2,11 +2,10 @@ package main
|
|||
|
||||
import (
|
||||
"context"
|
||||
"errors"
|
||||
"fmt"
|
||||
"time"
|
||||
|
||||
"github.com/simonvetter/modbus"
|
||||
"k8s.io/klog"
|
||||
)
|
||||
|
||||
func modbusValuesToFloat(v uint16) float32 {
|
||||
|
@ -39,7 +38,7 @@ func (d *daemon) modbusConnect() error {
|
|||
// There are currently two devices connected to the modbus.
|
||||
// The first one (slave 1) is a temperature/humidity sensor.
|
||||
// The second one (slave 2) is a PTA8D08 transmitter
|
||||
func (d *daemon) modbusUpdate() error {
|
||||
func (d *daemon) modbusUpdate() {
|
||||
var err error
|
||||
|
||||
// Switch to slave 1 (BTA1)
|
||||
|
@ -49,19 +48,16 @@ func (d *daemon) modbusUpdate() error {
|
|||
var registersBTA1 []uint16 // temperature, humidity
|
||||
registersBTA1, err = d.modbusClient.ReadRegisters(1, 2, modbus.INPUT_REGISTER)
|
||||
if err != nil {
|
||||
return err
|
||||
klog.Warningf("error while reading registers from BTA1 %v", err)
|
||||
} else if len(registersBTA1) != 2 {
|
||||
klog.Warningf("expected two registers from modbus slave 1, but got %d", len(registersBTA1))
|
||||
} else {
|
||||
d.mu.Lock()
|
||||
d.daemonState.tempSEM = modbusValuesToFloat(registersBTA1[0])
|
||||
d.daemonState.humiditySEM = modbusValuesToFloat(registersBTA1[1])
|
||||
d.mu.Unlock()
|
||||
}
|
||||
|
||||
if len(registersBTA1) != 2 {
|
||||
msg := fmt.Sprintf("Expected two registers from modbus slave 1, but got %d", len(registersBTA1))
|
||||
return errors.New(msg)
|
||||
}
|
||||
|
||||
d.mu.Lock()
|
||||
d.daemonState.tempSEM = modbusValuesToFloat(registersBTA1[0])
|
||||
d.daemonState.humiditySEM = modbusValuesToFloat(registersBTA1[1])
|
||||
d.mu.Unlock()
|
||||
|
||||
// Switch to slave 2 (KEC2)
|
||||
d.modbusClient.SetUnitId(2)
|
||||
|
||||
|
@ -72,19 +68,17 @@ func (d *daemon) modbusUpdate() error {
|
|||
var registersKEC2 []uint16 // temperatures dp
|
||||
registersKEC2, err = d.modbusClient.ReadRegisters(0, 3, modbus.HOLDING_REGISTER)
|
||||
if err != nil {
|
||||
return err
|
||||
klog.Warningf("error while reading registers from KEC2 %v", err)
|
||||
} else if len(registersKEC2) != 3 {
|
||||
klog.Warningf("expected three registers from modbus slave 2, but got %d", len(registersKEC2))
|
||||
} else {
|
||||
d.mu.Lock()
|
||||
d.daemonState.tempDPBottom = modbusValuesToFloat(registersKEC2[0])
|
||||
d.daemonState.tempDPInlet = modbusValuesToFloat(registersKEC2[1])
|
||||
d.daemonState.tempDPTop = modbusValuesToFloat(registersKEC2[2])
|
||||
d.mu.Unlock()
|
||||
}
|
||||
|
||||
if len(registersKEC2) != 3 {
|
||||
return fmt.Errorf("expected three registers from modbus slave 2, but got %d", len(registersKEC2))
|
||||
}
|
||||
|
||||
d.mu.Lock()
|
||||
d.daemonState.tempDPBottom = modbusValuesToFloat(registersKEC2[0])
|
||||
d.daemonState.tempDPInlet = modbusValuesToFloat(registersKEC2[1])
|
||||
d.daemonState.tempDPTop = modbusValuesToFloat(registersKEC2[2])
|
||||
d.mu.Unlock()
|
||||
|
||||
// Switch to slave 3 (KEC1)
|
||||
d.modbusClient.SetUnitId(3)
|
||||
|
||||
|
@ -94,17 +88,18 @@ func (d *daemon) modbusUpdate() error {
|
|||
|
||||
digitalInputRegisters, err = d.modbusClient.ReadRegisters(0x81, 8, modbus.HOLDING_REGISTER)
|
||||
if err != nil {
|
||||
return err
|
||||
}
|
||||
// Convert MODBUS words into bools
|
||||
for idx, value := range digitalInputRegisters {
|
||||
if value != 0 {
|
||||
digitalInputs[idx] = true
|
||||
} else {
|
||||
digitalInputs[idx] = false
|
||||
klog.Warningf("error while reading digital inputs from KEC1 %v", err)
|
||||
} else {
|
||||
// Convert MODBUS words into bools
|
||||
for idx, value := range digitalInputRegisters {
|
||||
if value != 0 {
|
||||
digitalInputs[idx] = true
|
||||
} else {
|
||||
digitalInputs[idx] = false
|
||||
}
|
||||
}
|
||||
// TODO: Input mapping goes here
|
||||
}
|
||||
// TODO: Input mapping goes here
|
||||
|
||||
// KFA1-KFA8
|
||||
var relayState [8]bool
|
||||
|
@ -140,10 +135,8 @@ func (d *daemon) modbusUpdate() error {
|
|||
|
||||
err = d.modbusClient.WriteRegisters(0x01, registerValuesKEC1[:])
|
||||
if err != nil {
|
||||
return err
|
||||
klog.Warningf("error while updating registers %v", err)
|
||||
}
|
||||
|
||||
return nil
|
||||
}
|
||||
|
||||
// Call modbusUpdate every 100 milliseconds
|
||||
|
|
Loading…
Reference in a new issue