jeol-t330a/succbone/succd/process.go

181 lines
4.6 KiB
Go
Raw Normal View History

package main
import (
"context"
"errors"
"fmt"
"sync"
"time"
"k8s.io/klog"
)
// daemon is the main service of the succdaemon.
type daemon struct {
// adcPirani is the adc implementation returning the voltage of the Pfeiffer
// Pirani gauge.
adcPirani adc
gpioDiffusionPump gpio
gpioRoughingPump gpio
gpioBtnPumpDown gpio
gpioBtnVent gpio
gpioBelowRough gpio
gpioBelowHigh gpio
// mu guards the state below.
mu sync.RWMutex
daemonState
}
// daemonState contains all the state of the daemon. A copy of it can be
// requested for consumers, eg. the web view.
type daemonState struct {
safety struct {
// failsafe mode is enabled when the pirani gauge appears to be
// disconnected, and is disabled only when an atmosphere is read.
failsafe bool
// highPressure mode is enabled when the pressure reading is above 1e-1
// mbar, locking out the diffusion pump from being enabled.
highPressure bool
}
piraniVolts100 ringbufferInput
2024-09-28 08:17:05 +00:00
piraniMbar100 pfeifferVoltsToMbar
piraniVolts3 ringbufferInput
2024-09-28 08:17:05 +00:00
piraniMbar3 pfeifferVoltsToMbar
rpOn bool
dpOn bool
vent momentaryOutput
pumpdown momentaryOutput
aboveRough thresholdOutput
aboveHigh thresholdOutput
}
func (d *daemonState) vacuumStatus() (rough, high bool) {
rough = !d.aboveRough.output
high = !d.aboveHigh.output
return
}
// process runs the pain acquisition and control loop of succd.
func (d *daemon) process(ctx context.Context) {
ticker := time.NewTicker(time.Millisecond * 100)
defer ticker.Stop()
for {
select {
case <-ticker.C:
if err := d.processOnce(ctx); err != nil {
if errors.Is(err, ctx.Err()) {
return
} else {
klog.Errorf("Processing error: %v", err)
time.Sleep(time.Second * 10)
}
}
case <-ctx.Done():
return
}
}
}
// processOnce runs the main loop step of succd.
func (d *daemon) processOnce(_ context.Context) error {
v, err := d.adcPirani.Read()
if err != nil {
return fmt.Errorf("when reading ADC: %w", err)
}
d.mu.Lock()
defer d.mu.Unlock()
// Process pirani ringbuffers.
d.piraniVolts3.process(v)
2024-09-28 08:17:05 +00:00
d.piraniMbar3.process(d.piraniVolts3.avg)
d.piraniVolts100.process(v)
2024-09-28 08:17:05 +00:00
d.piraniMbar100.process(d.piraniVolts100.avg)
d.pumpdown.process()
d.vent.process()
2024-09-28 08:17:05 +00:00
// Run safety checks based on small ringbuffer.
if d.piraniVolts3.saturated() {
mbar := d.piraniMbar3.mbar
if !d.safety.failsafe && mbar < 4e-6 {
// Unrealistic result, Pirani probe probably disconnected. Failsafe mode.
if !d.safety.failsafe {
d.safety.failsafe = true
2024-09-28 08:22:40 +00:00
klog.Errorf("SAFETY: Pirani probe seems disconnected; enabling failsafe mode")
}
}
2024-09-28 08:17:05 +00:00
if d.safety.failsafe && mbar > 1e2 {
d.safety.failsafe = false
2024-09-28 08:22:40 +00:00
klog.Infof("SAFETY: Pirani probe value (%s) is plausible again; quitting failsafe mode", formatMbar(mbar))
2024-09-28 08:17:05 +00:00
}
2024-09-28 08:17:05 +00:00
if !d.safety.highPressure && mbar >= 1e-1 {
d.safety.highPressure = true
2024-09-28 08:22:40 +00:00
klog.Warningf("SAFETY: Pressure is too high (%s mbar); enabling diffusion pump lockout", formatMbar(mbar))
}
2024-09-28 08:17:05 +00:00
if d.safety.highPressure && mbar < (1e-1)-(1e-2) {
d.safety.highPressure = false
2024-09-28 08:22:40 +00:00
klog.Infof("SAFETY: Pressure is low enough (%s mbar) for diffusion pump operation; quitting diffusion pump lockout", formatMbar(mbar))
2024-09-27 00:24:19 +00:00
}
2024-09-28 08:17:05 +00:00
} else {
d.safety.failsafe = true
d.safety.highPressure = true
2024-09-27 00:24:19 +00:00
}
2024-09-28 08:17:05 +00:00
// Control threhold/feedback values based on main pirani ringbuffer, failing
// safe if not enough data is present.
if d.piraniVolts100.saturated() {
mbar := d.piraniMbar100.mbar
d.aboveRough.process(float64(mbar))
d.aboveHigh.process(float64(mbar))
} else {
d.aboveRough.output = true
d.aboveHigh.output = true
}
// Apply safety overrides.
if d.safety.failsafe {
2024-09-27 00:24:19 +00:00
d.aboveRough.output = true
d.aboveHigh.output = true
d.dpOn = false
}
if d.safety.highPressure {
d.dpOn = false
2024-09-27 00:24:19 +00:00
}
// Update relay outputs.
for _, rel := range []struct {
name string
gpio gpio
// activeHigh means the relay is active high, ie. a true source will
// mean that NO/COM get connected, and a false source means that NC/COM
// get connected.
activeHigh bool
source bool
}{
{"rp", d.gpioRoughingPump, false, d.rpOn},
{"dp", d.gpioDiffusionPump, true, d.dpOn},
{"pumpdown", d.gpioBtnPumpDown, true, d.pumpdown.output},
{"vent", d.gpioBtnVent, true, d.vent.output},
{"rough", d.gpioBelowRough, false, d.aboveRough.output},
{"high", d.gpioBelowHigh, false, d.aboveHigh.output},
} {
val := rel.source
if rel.activeHigh {
// Invert because the relays go through logical inversion (ie. a
// GPIO false is a relay trigger).
val = !val
}
if err := rel.gpio.set(val); err != nil {
return fmt.Errorf("when outputting %s: %w", rel.name, err)
}
}
return nil
}